[21] X Li.X Fang.Z Yu.G Sheng.M Wu.Direct quantification of creatinine in human urine by using isotope dilution extractive electrospray ionization tandem mass spectrometry[J].《Analytica Chimica Acta》, 2012, 748(20):53-57
发布人:王国凤  发布时间:2017-03-14   动态浏览次数:106

Abstract:Urinary creatinine (CRE) is an important biomarker of renal function. Fast and accurate quantification of CRE in human urine is required by clinical research. By using isotope dilution extractive electrospray ionization tandem mass spectrometry (EESI–MS/MS) a high throughput method for direct and accurate quantification of urinary CRE was developed in this study. Under optimized conditions, the method detection limit was lower than 5002μg02L 611 . Over the concentration range investigated (0.05–1002mg02L 611 ), the calibration curve was obtained with satisfactory linearity ( R 2 02=020.9861), and the relative standard deviation (RSD) values for CRE and isotope-labeled CRE (CRE-d3) were 7.1–11.8% ( n 02=026) and 4.1–11.3% ( n 02=026), respectively. The isotope dilution EESI–MS/MS method was validated by analyzing six human urine samples, and the results were comparable with the conventional spectrophotometric method (based on the Jaffe reaction). Recoveries for individual urine samples were 85–111% and less than 0.302min was taken for each measurement, indicating that the present isotope dilution EESI–MS/MS method is a promising strategy for the fast and accurate quantification of urinary CRE in clinical laboratories.